TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE 6–25
Messages on the
Autosampler CE Display
The Autosampler CE display messages are
divided into two lists: fatal errors and nonfatal
errors. Some of the nonfatal errors are ignored,
allowing operation to continue, and some cause
the Autosampler CE to halt and wait for operator
intervention.
Fatal Error Messages
Fatal error messages indicate a condition that
must be corrected for operation to continue. The
Autosampler CE always stops operating when a
fatal error occurs.
Fatal Error
8155 U12
Problem: The confidence test has detected that
the Autosampler CE s scratch-pad
RAM or I/O (input/output) is not
properly storing what is written to it.
Solution: Turn off the Autosampler CE, and
power up again. If you still receive this
message, call a qualified service
representative.