Diagnostics
6–24 TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE
Error Messages
This section defines and offers solutions for
three types of Autosampler CE error messages:
(1) General error messages generated by the
controller about the Autosampler CE.
(2) Error messages displayed on the Autosam-
pler CE screen. These messages are
explained in alphabetical order starting on
page 6-25.
(3) Error messages generated by the Auto-
sampler CE and shown on the controller
screen.
General Error Messages
If you receive any general controller or instru-
ment error messages, please access the online
help provided with the Thermal Solutions/
Advantage software for detailed instructions,
and read the guidelines given here.
You can access online help by opening the Error
Log in
Thermal Solutions/Advantage
Instrument
Control and double-clicking on the error listed.
When an instrument error occurs, the AS run
sequence halts and waits for you to correct the
error. Once you have corrected the error, select
Control/Resume Motion from the Autosam-
pler CE menu to resume the AS run sequence. If
you cannot correct the error, select Control/
Reset or Control/Stop from the Autosampler
CE menu to stop the current run. Then select
Control/Start at and select the next run in the
run sequence to continue.