CE 6–23
Autosampler CE Test Functions
The screen displays:
Cycle A--B
Enter Point “B”
5. Use the motion keys to put the Autosampler
CE in the second position you wish to cycle
between; then press ENTER. The screen
displays:
Cycle A--B
Enter to Start
6. Press ENTER. The Autosampler CE
continuously cycles between points A and B.
Be ready to press the HALT key to prevent
possible collisions of the turret arm with the
enclosure, tray, or cell.
Once you are sure that the path the Auto-
sampler CE takes between points A and B
will not result in a collision, you can cycle
this test indefinitely.
7. Press ESCAPE/CLEAR to exit the Cycle
AB test.
When you have finished using the Autosampler
CE keypad, remember to close the keypad cover
and finger-tighten the thumbscrews before
starting a method.
t
CAUTION:
NOTE: