Diagnostics
6–22 TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE
t
CAUTION:
The Cycle A–B Test
The Cycle AB test cycles the Autosampler CE
between two positions of your choice to check
the Autosampler CE s motor operation.
The following test is provided primarily for TA
Instruments service use. Improper use of this
test can cause damage to the turret mecha-
nism. This test overrides the Autosampler CE’s
anticollision programming; collisions of the
turret arm with the enclosure, cell, or sample
tray are possible. NEVER leave this test operat-
ing unattended. Please contact an authorized
service representative for further information.
1. Access the Autosampler CE keypad by
loosening the captive thumbscrews on the
keypad cover and tilting the cover down.
2. Press the TEST key nine times. The
Autosampler CE screen displays:
Test
Cycle A--B
3. Press ENTER to select the test displayed on
the screen. The screen displays:
Cycle A--B
Enter Point “A”
4. Use the motion keys to put the Autosampler
CE in the first position you wish to cycle
between, then press ENTER.