TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE 6–27
Solution: Turn off the Autosampler CE, and
power up again. If you still receive this
message, call a qualified service
representative.
Fatal Error
RAM U5
Problem: The confidence test has detected that
the RAM storing data for the Auto-
sampler CE is not operating correctly.
Solution: Turn off the Autosampler CE, and
power up again. If you still receive this
message, call a qualified service
representative.
Fatal Error
Timer Interrupt
Problem: The confidence test has detected that
the master clock signal that times all
Autosampler CE functions is not
operating correctly.
Solution: Turn off the Autosampler CE, and
power up again. If you still receive this
message, call a qualified service
representative.
Error Messages