CE 6–29
Nonfatal Error Messages
Nonfatal error messages indicate a condition that
may require operator intervention to correct the
problem and allow the Autosampler CE to
continue.
Cal. Mem. Error
<position>
(calibration memory error)
Problem: The value for the indicated calibration
position has been erased.
Solution: Recalibrate the position displayed on
the screen (see Chapter 4).
Contact Error
Finger “X”
(X = A or B)
Problem: One of the gripper fingers has touched
the DSC cell. Motion is stopped when
this occurs, to prevent damage to the
cell or gripper fingers. All motion
keys except the RAISE key become
inoperable.
Solution: Press the RAISE key to clear the
gripper fingers from the cell. Then use
the other motion keys to reposition the
grippers before you retry the inter-
rupted task.
- OR -