TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE 6–31
Error Messages
proper commands; this indicates a
faulty part.
Solution: Turn off the Autosampler CE, and
power up again. If you still receive this
message, call a qualified service
representative.
Error
Holding unk. obj
(holding unknown object)
Problem: (1) The Autosampler CE was
switched to automatic mode
while it was holding an object or
was halted (with the Halt
Motion key) while the grippers
were closing on an object.
- OR -
(2) You manually gripped an object
with the motion keys, then
attempted to use the LOAD or
LIDS key.
Solution: (1) Switch to manual mode (see
Chapter 3), and release the
object with the RELEASE key.
Then return to automatic mode.
- OR -
(2) Release the object with the
RELEASE key; then retry your
command.