Diagnostics
6–32 TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE
Error
Keyboard shorted
Problem: One or more of the Autosampler CE
keys is shorted or was pressed during
the confidence test.
Solution: Turn off the Autosampler CE, and
power up again. If you still receive this
message, call a qualified service
representative.
Error
Lid #x dropped
Problem: The Autosampler CE has dropped a
lid (lid #1, #2, or #3) or had trouble
picking it up. This can be caused by
(1) incorrect calibration, (2) bent or
damaged gripper fingers, or (3) a
faulty Autosampler CE gripper
mechanism.
Solution: Try recalibrating the related
position(s); then repeat the task. If the
error occurs again, call a qualified
service representative.
Error
Lid #x missing
Problem: The Autosampler CE cannot find a lid
(lid #1, #2, or #3) either in its storage
position or on the DSC cell.