CE 6–33
Solution: Return the lid to its storage position.
When the Autosampler CE senses the
lid, it should return to normal operation.
If this does not work, test the lid
storage sensor with the TEST key
(see page 6-9).
If this error message appears when
the Autosampler CE is gripping a lid,
the gripper sensor may be malfunc-
tioning; test the gripper sensor with the
TEST key.
Error
Lid #x not strd.
(Lid #x not stored)
Problem: The Autosampler CE has been reset
with one of the Reset Sampler keys,
and one of the lids (lid #1, #2, or #3)
has not been returned to its storage
position.
Solution: Use tweezers to return the lid to its
storage position. If the Autosampler
CE is in manual mode, press ES-
CAPE/CLEAR to clear the error
message from the screen.
Error Messages