NSTRUMENTS
DSC A
UTOSAMPLER
CE 6–39
Error Messages
GPIB Error 54
non-num. AC type
(non-numeric AC type)
Problem: An undefined GPIB command was
sent from the controller to the Auto-
sampler CE.
Solution: Close the DSC Instrument Control
window, and turn off the Autosampler
CE. Then restart the Autosampler CE,
reopen DSC Instrument Control, and
retry your operation.
GPIB Error 55
unknown AC type
Problem: An undefined GPIB command was
sent from the controller to the Auto-
sampler CE.
Solution: Close the DSC Instrument Control
window, and turn off the Autosampler
CE. Then restart the Autosampler CE,
reopen DSC Instrument Control, and
retry your operation.
GPIB Error 56
bad sequence num
(bad sequence number)
Problem: The sequence number of a GPIB
command sent to the Autosampler CE
contains a non-numeric character.