Solution: Close the DSC Instrument Control
window, and turn off the Autosampler
CE. Then restart the Autosampler CE,
reopen DSC Instrument Control, and
retry your operation.
GPIB Error 57
no sequence num.
(no sequence number)
Problem: A GPIB command sent to the Auto-
sampler CE does not contain a se-
quence number.
Solution: Close the DSC Instrument Control
window and turn off the Autosampler
CE. Then restart the Autosampler CE,
reopen DSC Instrument Control, and
retry your operation.
GPIB Error 58
large sequen num
(large sequence number)
Problem: The sequence number of a GPIB
command sent to the Autosampler CE
is too large.
Solution: Close the DSC Instrument Control
window, and turn off the Autosampler
CE. Then restart the Autosampler CE,
reopen DSC Instrument Control, and
retry your operation.