TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE 4–19
23. When you are satisfied that the pan is
properly positioned, press the CALIBRATE
key repeatedly until you see the correct cell
pan position on the display:
Calibrate/Verify
Sample Pos.
24. Press ENTER. The screen displays:
Verify (1) or
Store (2) ? 1
25. Select option 2 (Store) to store the calibra-
tion position (press the 2 key, then press
ENTER). The screen displays:
Sample Pos.
Calibrated
26. Release the pan (press the RELEASE key).
Press RESET SAMPLER and ENTER to
bring the Autosampler CE to the home
position. Press the CALIBRATE key until
you see the correct cell lid position on the
display:
Calibrate/Verify
Sample Pos.
27. Press ENTER. The screen displays:
Verify (1) or
Store (2) ? 1
Performing the Calibration