Calibrating the Autosampler CE
4–20 TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE
NOTE:
28. Press ENTER to select default option 1
(Verify).
29. When the turret stops moving, close and
open the grippers again and check your
calibration against the criteria in Figure 4.2.
Remember to release the pan and raise the
turret out of the cell with the RAISE key.
30. Use the LOAD key to remove the pans from
the cell.
31. Press ESCAPE/CLEAR to end the calibra-
tion procedure. Go on to the section entitled
Checking Your Calibration (page 4-29).
When you have finished using the Autosampler
CE keypad, remember to close the keypad cover
and finger-tighten the thumbscrews before
starting a method.