EasyManuals Logo

TA Instruments DSC Autosampler CE User Manual

TA Instruments DSC Autosampler CE
190 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #91 background imageLoading...
Page #91 background image
TA I
NSTRUMENTS
DSC A
UTOSAMPLER
CE 4–23
Performing the Calibration
The screen displays:
Lid #1 Storage
Calibrated
8. Release the lid (press the RELEASE key).
Press RESET SAMPLER and ENTER to
bring the Autosampler CE to the home
position. Press the CALIBRATE key until
the correct cell lid position is on the display:
Calibrate/Verify
Lid #1 Storage
9. Press ENTER. The screen displays:
Verify (1) or
Store (2) ? 1
10. Press ENTER to select default option 1
(Verify).
11. When the turret stops moving, close and
open the grippers again and check your
calibration against the criteria in Figure 4.3.
Remember to release the lid. If the position
is correctly calibrated, go on to step 12. If it
is not, repeat steps 4 through 11 with lid #1.
12. Repeat steps 4 through 11 to calibrate the
storage positions for lids 2 and 3 (the me-
dium-sized and large lids). The screen
displays in steps 5, 7, and 8 are the same
except for the lid numbers, as shown on the
next page.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TA Instruments DSC Autosampler CE and is the answer not in the manual?

TA Instruments DSC Autosampler CE Specifications

General IconGeneral
BrandTA Instruments
ModelDSC Autosampler CE
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals