Performing the Calibration
The screen displays:
Lid #1 Storage
Calibrated
8. Release the lid (press the RELEASE key).
Press RESET SAMPLER and ENTER to
bring the Autosampler CE to the home
position. Press the CALIBRATE key until
the correct cell lid position is on the display:
Calibrate/Verify
Lid #1 Storage
9. Press ENTER. The screen displays:
Verify (1) or
Store (2) ? 1
10. Press ENTER to select default option 1
(Verify).
11. When the turret stops moving, close and
open the grippers again and check your
calibration against the criteria in Figure 4.3.
Remember to release the lid. If the position
is correctly calibrated, go on to step 12. If it
is not, repeat steps 4 through 11 with lid #1.
12. Repeat steps 4 through 11 to calibrate the
storage positions for lids 2 and 3 (the me-
dium-sized and large lids). The screen
displays in steps 5, 7, and 8 are the same
except for the lid numbers, as shown on the
next page.