:DATE
(:CALibration:PROTected:DATE)
Purpose To program the calibration date.
Format :cal:prot:date <year>, <month>, <day>
Parameters <year> = 1996 to 2095
<month> = 1 to 12
<day> = 1 to 31
Query :cal:prot:date?
Response <year> , <month> , <day>
Description The :DATE command allows you to store the calibration date in instrument
EEROM for future reference. You can read back the date from the instrument
by using the :DATE? query or by using the front panel CAL menu.
Note The year, month, and day parameters must be delimited by commas.
Example :CAL:PROT:DATE 1996,11,20 Send cal date (11/20/96).
:NDUE
(:CALibration:PROTected:NDUE)
Purpose To send the next calibration due date to the instrument.
Format :cal:prot:ndue <year>, <month>, <day>
Parameters <year> = 1996 to 2095
<month> = 1 to 12
<day> = 1 to 31
Query :cal:prot:ndue?
Response <year>, <month>, <day>
Description The :NDUE command allows you to store the date when calibration is next
due in instrument memory. You can read back the next due date by using the
:NDUE? query or by using the front panel CAL menu.
Note The next due date parameters must be delimited by commas.
Example :CAL:PROT:NDUE 1997,11,20 Send due date (11/20/97).
Command Reference B-5
2410-902-01.book Page 5 Monday, November 7, 2005 2:49 PM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2410 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2410 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals