Performance Verification 1-7
Performing the verification test procedures
Test summary
• DC voltage output accuracy
• DC voltage measurement accuracy
• DC current output accuracy
• DC current measurement accuracy
• Resistance measurement accuracy
If the Model 2410 is not within specifications and not under warranty, see the calibration
procedures in Section 2 for information on calibrating the unit.
Test considerations
When performing the verification procedures:
• Be sure to restore factory front panel defaults as outlined above.
• Make sure the test equipment is properly warmed up and connected to the Model 2410
INPUT/OUTPUT
jacks. Also be sure that the front panel jacks are selected with the
TERMINALS key.
• Make sure the Model 2410 is set to the correct source range.
• Be sure the Model 2410 output is turned on before making measurements.
• Be sure the test equipment is set up for the proper function and range.
• Allow the Model 2410 output signal to settle before making a measurement.
• Do not connect test equipment to the Model 2410 through a scanner, multiplexer, or
o
t
he
r switching equipment.
WARNING The
max
imum common-mode voltage (voltage between LO and chassis
CAUTION
g
round) is 250V peak. Exceeding this value may cause a breakdown in
insulation, creating a shock hazard.
The max
imum voltage between INPUT/OUTPUT HI and LO or 4-WIRE
SENSE HI and LO is 1100V peak. The maximum voltage between INPUT/
OUTPUT HI and 4-WIRE SENSE HI or between INPUT/OUTPUT LO
and 4-WIRE SENSE LO is 5V. Exceeding these voltage values may result
in instrument damage.
2410-902-01.book Page 7 Monday, November 7, 2005 2:49 PM