Note: Actual allowable parameter ranges are ±50% of full range for zero parameters, 75%
of full scale to 150% of full scale for full-range parameters, but ranges listed above should
be observed for optimum calibration.
Table B-2
Recommended :CALibration:PROTected:SENSe parameter ranges
Sense
rang
e
First parameter
(zero)
Second parameter
(negative full scale)
Third parameter
(positive full scale)
0.2V
2V
20V
1000V
-0.002 to +0.002
-0.02 to +0.02
-0.2 to +0.2
-2 to +2
-0.18 to -0.22
-1.8 to -2.2
-18 to -22
-800 to -1100
+0.18 to +0.22
+1.8 to +2.2
+18 to +22
+800 to +1100
1µA
10µA
100µA
1mA
20mA
100mA
1A
-1E-8 to +1E-8
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
-0.9E-6 to -1.1E-6
-9E-6 to -11E-6
-90E-6 to -110E-6
-0.9E-3 to -1.1E-3
-18E-3 to -22E-3
-90E-3 to -110E-3
-0.9 to -1.1
+0.9E-6 to +1.1E-6
+9E-6 to +11E-6
+90E-6 to +110E-6
+0.9E-3 to +1.1E-3
+18E-3 to +22E-3
+90E-3 to +110E-3
+0.9 to +1.1
Command Reference B-7
2410-902-01.book Page 7 Monday, November 7, 2005 2:49 PM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2410 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2410 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals