Power supply checks
Power supply problems can be checked out using Table 4-2. See principles of operation for
circuit theory on the power supply.
Table 4-2
Power supply checks
Step Item/component Required condition Remarks
1
2
3
4
5
6
7
8
9
10
11
Line fuse
Line power
CR500 cathode
CR503 anode
CR504 cathode
CR505 anode
TP506
TP507
TP508
TP509
TP510
Check continuity.
Plugged into live
receptacle, power on.
+1200V ±10%
-1200V ±10%
+4
2V ±10%
-42V ±10%
+70
V ±10%
+15V ±5%
-15V ±10%
-70V ±10%
+5V ±5%
Remove to check.
Check for correct power-up
seq
uenc
e.
Referenced to TP501.
Referenced to TP501.
Referenced to TP501.
Referenced to TP501.
+70VF, referenced to TP500.
+15VF, referenced to TP500.
-15VF, referenced to TP500.
-70VF, referenced to TP500.
+5VF, referenced to TP500.
4-14 Troubleshooting
2410-902-01.book Page 14 Monday, November 7, 2005 2:49 PM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2410 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2410 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals