Single-range calibration
Normally, the complete calibration
procedure should be performed to ensure that the entire
instrument is properly calibrated. In some instances, however, you may want to calibrate only
certain ranges. To do so, complete the entire procedure only for the ranges to be calibrated.
Keep in mind, however, that you must complete all parameter steps for each source or sense
range. Also be sure to set calibration dates and save calibration after calibrating the desired
ranges.
2-24 Calibration
2410-902-01.book Page 24 Monday, November 7, 2005 2:49 PM