Step 4: Program calibration dates
Use following commands to set the calibration date and calibration due date:
:CAL:PROT:DATE <year>, <month>, <day> (Calibration date)
:CAL:PROT:NDUE <year>, <month>, <day> (Next calibration due date)
Note that the year, month, and day must be separated by commas. The allowable range for
the year is from 1996 to 2095, the month is from 1 to 12, and the day is from 1 to 31.
Step 5: Save calibration constants
Calibration is now complete, so you can store the calibration constants in EEROM by sending
the following command:
:CAL:PROT:SAVE
NOTE Calibration will be temporary unless you send the SAVE command. Also, calibration
d
ata will not be saved if (1) calibration is locked, (2) invalid data exists, or (3) all
steps were not completed.
Step 6: Lock out calibration
To lock out further calibration, send the following command after completing the calibration
procedure:
:CAL:PROT:LOCK
Calibration 2-23
2410-902-01.book Page 23 Monday, November 7, 2005 2:49 PM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2410 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2410 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals