1-4 Performance Verification
Recommended test equipment
Table 1-1 summarizes recommended v
erification equipment. You can use alternate
equipment if that equipment has specifications at least as good as those listed in Table 1-1. Be
aware, however, that test equipment uncertainty will add to the uncertainty of each
measurement. Generally, test equipment uncertainty should be at least four times better than
corresponding Model 2410 specifications. Table 1-1 lists the uncertainties of the
recommended test equipment including the maximum allowable uncertainty of alternate test
equipment shown in parentheses.
Table 1-1
Recommended verification equipment
Description
Manufacturer/
Mod
el Accuracy
Digital multimeter Hewlett Packard
H
P
3458A
DC voltage*
DC current*
1V:
10V:
100V:
1000V:
1µA:
10μA:
100μA:
1mA:
100mA:
1A:
±5.6ppm
±4.3ppm
±6.3ppm
±6.1ppm
±55ppm
±25ppm
±23ppm
±20ppm
±35ppm
±110ppm
Resistance calibrator Fluke 5450A Resistance** 19Ω:
190Ω:
1.9kΩ:
19kΩ:
190kΩ:
1.9MΩ:
19MΩ:
100MΩ:
±23ppm340ppm)
±10.5ppm (±370ppm)
±8ppm (±265ppm)
±7.5ppm (±250ppm)
±8.5ppm (±260ppm)
±11.5ppm (±365ppm)
±30ppm315ppm)
±120ppm (±1700ppm)
** 90-day full-scale accuracy specifications of ranges required for various measurement points.
** 90-day specifications of nominal resistance values shown. Maximum uncertainty of alternate test
equ
ipment is shown in parentheses.
2410-902-01.book Page 4 Monday, November 7, 2005 2:49 PM

Table of Contents

Related product manuals