Note: Actual allowable parameter ranges are ±50% of full range for zero parameters, 75% of full scale to 150% of
full scale for full-range parameters, but ranges listed above should be observed for optimum calibration.
Table B-3
Recommended :CALibration:PROTected:SOURce parameter ranges
Source
ra
nge
First parameter
(negative full scale)
Second
pa
rameter
(negative zero)
Third parameter
(po
sitive full scale)
Fourth parameter
(positive zero)
0.2V
2V
20V
1000V
-0.18 to -0.22
-1.8 to -2.2
-18 to -22
-800 to -1100
-0.002 to +0.002
-0.02 to +0.02
-0.2 to +0.2
-2 to +2
+0.18 to +0.22
+1.8 to +2.2
+18 to +22
+800 to +1100
-0.002 to +0.002
-0.02 to +0.02
-0.2 to +0.2
-2 to +2
1µA
10µA
100µA
1mA
20mA
100mA
1A
-0.9E-6 to -1.1E-6
-9E-6 to -11E-6
-90E-6 to -110E-6
-0.9E-3 to -1.1E-3
-18E-3 to -22E-3
-90E-3 to -110E-3
-0.9 to -1.1
-1E-8 to +1E-8
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
+0.9E-6 to +1.1E-6
+9E-6 to +11E-6
+90E-6 to +110E-6
+0.9E-3 to +1.1E-3
+18E-3 to +22E-3
+90E-3 to +110E-3
+0.9 to +1.1
-1E-8 to +1E-8
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
Command Reference B-9