:DATA?
(:CALibration:PROTected:SENSe:DATA?)
(:CALibration:PROTected:SOURce:DATA?)
Purpose To request the calibration constants for the active range.
Format :cal:prot:sens:data?
:cal:prot:sour:data?
Response Four comma-separated ASCII floating-point constants.
Description The :CAL:PROT:SENS:DATA? and :CAL:PROT:SOUR:DATA? queries
request the calibration constants for the active range of the sense and source
functions respectively. The four returned constants are in ASCII floating-
point format delimited by commas.
Note To request the appropriate constants, (1) select the source or sense function,
(2
) cho
ose the desired range, and (3) send the desired :DATA? query.
Example :SENS:FUNC 'VOLT:DC' Select DC sense function.
:SENS:VOLT:DC:RANGE 0.2 Choose 200mV range.
:CAL:PROT:SENS:DATA? Request cal constants.
B-10 Command Reference
2410-902-01.book Page 10 Monday, November 7, 2005 2:49 PM