Performance Verification 1-3
Verification test requirements
Be sure that you perform the verification tests:
• Under the proper environmental conditions.
• After the specified warm-up period.
• Using the correct line voltage.
• Using the proper test equipment.
• Using the specified output signal and reading limits.
Environmental conditions
Conduct your performance verification procedures in a test environment with:
• An ambient temperature of 18-28
°C.
• A relative humidity of less than 70% unless otherwise noted.
Warm-up period
Allow the Model 2410 to
warm up for at least one hour before conducting the verification
procedures. If the instrument has been subjected to temperature extremes (those outside the
ranges stated above), allow additional time for the instrument’s internal temperature to stabilize.
Typically, allow one extra hour to stabilize a unit that is 10°C outside the specified
temperature range.
Also, allow
the test equipment to warm up for the minimum time specified by the manufacturer.
Line power
The Model 2410 requires a line voltage of 100 to 240V and a line frequency of 50 Hz to
60 Hz. Verification tests must be performed within this range.