Table 2-4
Front panel current calibration
Source
range*
So
urce
current
Multimeter current
reading**
1µA
10µA
100µA
1mA
20mA
100mA
1A
+1.0000µA
+0.0000µA
-1.0000µA
-0.0000µA
+10.000µA
+00.000µA
-10.000µA
-00.000µA
+100.00µA
+000.00µA
-100.00µA
-000.00µA
+1.0000mA
+0.0000mA
-1.0000mA
-0.0000mA
+20.000mA
+00.000mA
-20.000mA
-00.000mA
+100.00mA
+000.00mA
-100.00mA
-000.00mA
+1.0000A
+0.0000A
-1.0000A
-0.0000A
____________ µA
____________ µA
____________ µA
____________ µA
____________ µA
____________ µA
____________ µA
____________ µA
____________ µA
____________ µA
____________ µA
____________ µA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
_____________ A
_____________ A
_____________ A
_____________ A
**Use EDIT and RANGE keys to select source range.
**Multimeter reading used in corresponding calibration
step. See procedure.
2-14 Calibration