Analog circuitry checks
Table 4-4 summarizes analog circuitry checks.
** Measured with respect to FCOM (TP500).
** Measured with respect to OCOM (TP501).
Table 4-4
Analog circuitry checks
Step
Item/
co
mponent Required co
ndition Remarks
1*
2*
3*
4*
5*
6**
7**
8**
9*
TP200
TP201
TP202
TP203
TP213
TP218
TP219
TP214
TP232
>1100V voltage protection
SOURCE +10V
SOURCE +10V (SVMI)
SOURCE +10V
SOURCE +10V
OUTPUT COM
OUTPUT COM
SVMI, OUTPUT ON, 20V
Bench defaults
-13V ±1V
-5V ±.5V
-10V ±1V
-10.5V ±1V
0V ±.1V
7V ±.7V
7V ±.7V
20V ±.5V
6.4V ±.6V
4-16 Troubleshooting