Model 880 DEPOSITION CONTROLLER
y
SECTION 6.XX
e page 209 of 276 ^
Parameter 10 Clock Time
access value: RTHST real time hours:minutes set param access id.
Value Range 00:00 - 23:59
Units Hours:Minutes
______________________________________________________________________
Parameter 11 Clock Date
access value: RTCLS real time date set MMDDYY 010100 .. 123199. param access id.
Value Range 010189 To 123188
Units Month Day Year
These 2 parameters program the clock of the Model 880. These are used in the process accounting mode.
The format for the time is 24 hour time. One O'clock in the afternoon is 13:00. The date is programmed in
month-day-year format without any separators. You need to enter any zeros in the day (March 5th is '05',
not just '5'). The range of dates is from 01/01/1989 through 12/31/2088
______________________________________________________________________
Parameter 12 LCD Contrast/Bias Control
access value: CRTL crt control, 1 of 3, 0=dim, 1=medium, 2=bright.
Value Range 1,2,3
Units LOW, MEDIUM, HIGH
This parameter adjusts the contrast/bias of the LCD.
______________________________________________________________________
Parameter 13 Beeper Control Enabling The Beeper
access value: BEEP beep, on/off type 0=off, 1=on.
Value Range 0 To 1
Units 0 is Off, 1 is On
This parameter controls if the beeper of the Model 880 is turned On or Off.
______________________________________________________________________
Parameter 14 Baud Rate
access value: BAUD baud , 1 of 4, 0=300, 1=1200, 2=2400, 3=9600. Takes effect immediately. This
will most likely break current communication
Value Range 0 To 3
Units 0 is 300, 1 is 1200, 2 is 2400, And 3 is 9600 Baud
This parameter controls the baud rate of the Model 880. It takes effect immediately. This likely means
future communications will fail until the host is set to the new value.
______________________________________________________________________
Parameter 15 Protocol Selection Communications Protocol Selection
access value: PROT protocol, selection, SYCON(0)/SECS(1), requires reset.
Value Range 0,1,2
Units 0,1 is Sycon Protocol, 2 is ASCII Protocol
This parameter selects which RS-232 protocol will be used, either Sycon or simple ASCII. It takes effect
only during a power up condition.
______________________________________________________________________
Parameter 16 NA access value: OBSO Was secs address.
______________________________________________________________________
Parameter 17 NA access value: OBSO Was link mode.
______________________________________________________________________
Parameter 18 NA access value: OBSO Was ieee-488 address.
______________________________________________________________________