3 Specifications
19
Parameter Accuracy Response
Temperature ± 0.5 °C (0.0...100.0 °C)
±0.5 % of mv (rest of range)
probe
dependent
Efficiency net - -
Flue gas loss - -
CO2a, through
0632 1240
±75 ppm + 3 % of mv
(0...5000 ppm)
±150 ppm + 5 % of mv
(5001...10000 ppm)
approx. 35 s
(t90)
Gas leak testing
with 0632 3330
- < 2 s (t90)
3.2.6. Other device data
Flue gas analyzer
Storage / and
transport
temperature
-20...50 °C
Operating
temperature
-5...45 °C
Ambient humidity 0…90 % rH, not condensing
Power supply Rech. batt: 3.7 V / 2.6 Ah
Mains unit: 6 V/1.2 A
Protection class IP40
Weight 600 g (without rech. batt.)
Dimensions 270 x 90 x 65 mm
Display Graphic color display, 240 x 320 pixels
Flue gas
overpressure
max. 50 mbar
Negative pressure max. 80 mbar
Gas leak testing
probe
visual indication (LED)
audible indication by buzzer