6 Using the product
72
- The stability time will run. After this the measuring phase starts
automatically.
> End stability time and measurement early: [Next].
- The readings and Result Pretest are displayed when
measurement has been completed.
6. [Edit] → Select test result: [▲], [▼] → [OK].
6.2.15.4. Leak detection
In gas leak detection no measurement is made, but a gas detection
is performed.
✓ A gas leak probe (0632 3330) must be connected.
You must also refer to the documentation that comes with
the gas leak probe.
Calling up the function:
> Leak Detection→ [OK].
Performing the detection:
> Set the gas type to be detected and perform gas leak testing by
following the instructions described in the documentation of the
gas leak probe.
- The gas concentration is displayed, if it exceeds the alarm limit
an alarm signal will sound.
Options:
> [Options] → Save: the readings are saved in a protocol.
> [Options] → Show Graphic: The meas. views are displayed in
form of a line graph.
> [Options] → Alarm Limit: (This function is not available during
a measurement) Adjust alarm thresholds.
> [Options] → Alarm Signal: (This function is not available
during a measurement) Disable/enable alarm signal.
> [Options] → Zeroing Probe: Perform zeroing.
> [Options] → Sensor detection: newly inserted sensor is
detected.