8 Tips and assistance
85
8.3. Updating the instrument software
Under www.testo.com/download-center you can download the
current instrument software (Firmware) for testo 330 (registration
required).
> Unplug the mains unit and switch off the testo 330.
1. Hold [▲] depressed.
2. Plug in the mains unit, keep [▲] depressed.
- The display shows Firmware update along the bottom edge.
3. Release [▲] .
4. Insert the connecting cable (0449 0047) into the USB-port on
the instrument, then connect it to the PC.
- Your PC recognizes the testo 330 as a removable medium.
5. Copy the new file (ap330rel.bin) to the detected removable
medium.
- In the display the status bar progresses from left to right. This
process may take a few minutes.
6. Disconnect the connecting cable from the testo 330.
- After updating of the instrument software (Firmware) has been
completed the system will automatically reboot and is ready for
use.

Table of Contents

Other manuals for TESTO 330

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TESTO 330 and is the answer not in the manual?

TESTO 330 Specifications

General IconGeneral
Protection classIP40
O2 measurement range0 to 21 % O2
CO measurement range0 to 4000 ppm
Measurement parametersO2, CO, CO2
Resolution0.1 % O2
Storage temperature-20 to +50 °C