Thermo Scientific 310 - Setting the CO2 Setpoint

Thermo Scientific 310
86 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
All T/C CO2 cells are calibrated at the factory at 37°C, high humidity, and
10% CO
2. Therefore, if a temperature setpoint of 37°C has been entered,
the humidity pan has been filled and the CO
2 control is to run between 0-
10% with a T/C CO
2 sensor, the CO2 setpoint may be entered
immediately. Otherwise, it is important to allow the unit 12 hours to
stabilize at the temperature setpoint before entering the CO
2 setpoint.
All models of the incubator have a CO
2 setpoint range of 0.0% to 20.0%.
The incubator is shipped from the factory with a CO
2 setpoint of 0.0%.
At this setting, all CO
2 control and alarms are turned off. To change the
CO
2 setpoint:
1. Press the Mode key until the Set indicator lights.
2. Press the right arrow until “CO2 XX.X” is displayed in the message
center.
3. Press the up/down arrows until the desired CO
2 setpoint is displayed.
4. Press Enter to save the setpoint.
5. Press the Mode key until the Run indicator lights to go to Run mode
or press the right/left arrow keys to go to next/previous parameter.
Model 310 Series DH Incubator 1-13Thermo Scientific
Section 1
Installation and Start-Up
Setting the CO2 Setpoint