Model 310 Series DH Incubator 3-3Thermo Scientific
Section 3
Configuration
The low CO2 alarm limit (tracking alarm) is the deviation from the CO2
setpoint which will cause a low CO2 alarm. The setpoint is variable from
0.5% CO
2 below setpoint to 5.0% CO2 below setpoint. The factory
setting is 1.0% CO
2 below setpoint. A minus sign (-) in the display
indicates that the alarm setting is below the setpoint.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until CO2 Lo -X.X is displayed in the message
center.
3. Press up/down arrow to change the low CO
2 alarm limit.
4. Press Enter to save the low CO
2 alarm limit.
5. Press the Mode key to return to Run mode or right/left to go to
next/previous parameter.
The high CO
2 alarm limit (tracking alarm) is the deviation from the CO2
setpoint which will cause a high CO2 alarm. The setpoint is variable from
0.5% CO
2 above setpoint to 5.0% CO2 above setpoint. The factory
setting is 1.0% CO
2 above setpoint.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until CO2 Hi X.X is displayed in the message
center.
3. Press up/down arrow to change the high CO
2 alarm limit.
4. Press Enter to save the high CO
2 alarm limit.
5. Press the Mode key to return to run mode or right/left to go to
next/previous parameter.
Set a Low CO2 Alarm
Limit
Setting a High CO
2
Alarm Limit