Table 4-1. Alarms and Descriptions
Model 310 Series DH Incubator 4-1Thermo Scientific
Section 4 Alarms
The Model 310 Series incubator alarm system is shown in the table below.
When an alarm is active, the message appears in the LED message center.
Pressing Silence disables the audible alarm for the ringback period.
However, the visual alarm continues until the incubator returns to a
normal condition. The alarms are momentary alarms only. When an alarm
condition occurs and then returns to normal, the incubator automatically
clears the alarm condition and the message center.
When multiple alarm conditions occur, active messages are displayed in
the message center one at a time, updating at 5 second intervals. Pressing
Silence during multiple alarms causes all active alarms to be silenced and to
ring back in 15 minutes.
The temperature alarms are disabled when the Temp set point is 10°C.
The CO
2 alarms are disabled when the CO2 set point is 0.0%.
Description Message Delay Ringback Relay
No alarm condition exists SYSTEM OK ----- ----- -----
CO2 System Auto Calibrating CO2 AUTO CAL ----- ----- -----
Temp > Otemp Set point SYS IN OTEMP 0 min. 15 min. Yes
Air Temp Sensor Fault (See Section 4) TMP SNSR ERR 0 min. 15 min. No
Temperature Controller Failure (See Section 4) TMP CTRL ERR 0 min. 15 min. YES
CO2 Sensor Fault (See Section 4) CO2 SNSR ERR 0 min. 15 min. No
CO2 Auto Zero Fault (IR option only, See Section 4) IR AUTOZ ERR 0 min. 15 min. No
CO2 Sensor Cannot be Calibrated (IR option only, See Section 4) REPL IR SNSR 0 min. 15 min. No
Inner Door is Open DOOR OPEN 15 min. 15 min. No
CO2 is higher than CO2 High Tracking Alarm CO2 IS HIGH 15 min. 15 min. Programmable
CO2 is lower than CO2 Low Tracking Alarm CO2 IS LOW 15 min. 15 min. Programmable
TEMP is lower TEMP Low Tracking Alarm TEMP IS LOW 15 min. 15 min. Programmable
RH is lower than RH Low Limit Alarm (RH option) RH IS LOW 30 min. 15 min. Programmable
Tank 1 is low, switch to Tank 2 (Gas Guard only) TANK1 LOW 0 min. ---- No
Tank 2 is low, switch to Tank 1 (Gas Guard only) TANK2 LOW 0 min. ---- No
Both tanks are low (Gas Guard only) TANK 1 and 2 LOW 0 min. 15 min. No
- All alarm delays and ringback times are ±30 seconds -