Section 7 Specifications
* Specifications are based on nominal voltages of 115V or 230V in ambient of 22°C
to 25°C.
Temperature
Control . . . .±0.1°C Microprocessor PID Control
Setpoint . . . . . . . . . . .Digital - Touch pad, 0.1°C
Range . . . . . . . . . . . .+5°C above ambient to 50°C
Uniformity . . . . . . . . . . . . . . . . . .±0.3°C @ +37°C
Tracking alarm . . User programmable (low) indicator
Overtemp . .Tracking, user programmable, action, and indicator
Display . . . . . . . . .Digital, LED, 0.1°C increments
Temperature Safety
Type . . Extreme temperature safety, action, and indicator
Sensor . . Thermostat, independent of temperature control system
Indicator . . Message center, audible and visual alarms
Relative Humidity
Control . . . . .Humidity pan - natural vaporization
Humidity with pan . . . . . . . . . . .95% RH at 37°C
Display . . . . . . . . . . . . .Optional in 1% increments
Alarm . . . . . . .Low RH with optional RH monitor
CO
2
Control . . . . .±0.1% microprocessor PID control
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . .T/C or IR
Readability . . . . . . . . . . . . . . . . . . . . . . . . . . .0.1%
Range . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 to 20%
Inlet pressure . . . . . . . . . . . . . . . . .15 psig, ±5 psig
Display . . . . . . . . . .Digital LED, 0.1% increments
Model 310 Series DH Incubator 7-1Thermo Scientific