On units equipped with the Gas Guard option, a primary tank can be
selected. The primary tank will be either Tank 1 or 2. The factory setting
is Tank1.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until Tnk Sel X is displayed in the message
center.
3. Press up/down arrow to toggle the setting between 1 and 2.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run mode or right/left to go to
next/previous parameter.
On units equipped with the Gas Guard option, the Gas Guard system
may be turned ON, or OFF if not in use. The factory setting is ON.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until Gas Grd XX is displayed in the message
center.
3. Press up/down arrow to toggle the setting ON/OFF.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run mode or right/left to go to
next/previous parameter.