The low RH alarm can be programmed to trip the remote alarm contacts.
A setting of ON will cause this, a setting of OFF will not allow the RH
alarm to trip the contacts. The factory setting is ON.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until RH Rly XXX is displayed in the message
center.
3. Press up/down arrow to toggle the setting ON/OFF.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run mode or right/left to go to
next/previous parameter.
On units that are equipped with the RH option, the upper seven segment
display on the control panel can be configured to display Temp
continuously, RH continuously, or toggle between Temp and RH. If the
units does not have RH, the upper display will always display temperature.
If temperature is set to ON and the RH is set OFF, temperature will be
displayed continuously. If temperature is set to OFF and RH is set to
ON, RH will be displayed continuously. If both are turned ON, the
display will toggle between the two. The factory setting will default to
toggle mode if the RH option is present.
1. Press the Mode key until the Config indicator lights.
2. Press the right arrow until Disp Tmp XXX or Disp RH XXX is
displayed in the message center.
3. Press up/down arrow to toggle the setting ON/OFF.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run mode or right/left to go to
next/previous parameter.
3-6 Model 310 Series DH Incubator Thermo Scientific
Section 3
Configuration
Enable Temp/RH to be
Displayed
Enable RH Alarms to
Trip Contacts