16
Operating Instruction TNI soft Flow 50 homecare system
A
B
3.1.2)Assemblingandinsertingthehumidiercliniccomplete
• Thehumidiercliniccompletecanbeusedalternativelyandcontains:
o “Humidierrackclinic”
o “Clear-Guard3lter”
o “Humidicationchamberauto-ll”
o “Airlifthumidierclinic”
• Removethe“humidierrackclinic”andthe“humidierclinic-
hygienicset”fromthepackagingandassemblethemaccordingto
theoperatinginstructions.
• Pleasereferadditionallytothedescriptiveimagesequence.
• Pushthe“humidiercliniccomplete”fromfrontintothedeviceand
closethefrontofthehousing.
• Fixasterilewaterbagtotherespectivehoseofthe“humidication
chamberauto-ll”(A)
• orusethe“humidicationchamberwithreservoir”(B).
WARNING
Hygiene
• Use authorized parts only.
• Make sure that the hygiene regulations are met.
• Use originally packed and unexpired parts only.
• Do not apply already used disposables, e.g. “humidication chamber
auto-ll”, “Clear-Guard 3 lter”, “Airlift humidier clinic” etc.
• Use sterile water in clinical treatment.
SAFETY NOTE
In order to ensure an optimal treatment,
• do not use the “humidication chamber auto ll” after it had been
dropped or run dry, which will trigger the alarm “rell water”.
• remove the “humidier clinic complete” BEFORE transporting, tilting or
moving the device.