EasyManua.ls Logo

TNI softFlow 50 - 1) This Manual; 2) Appropriate Application; 3) Correct Use

TNI softFlow 50
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7
Operating Instruction TNI soft Flow 50 homecare system
1.3.1)Thismanual
• Exactadherencetothefollowinginstructionsisaprerequisiteforthesafeandintendedoperation
ofthe
TNI soft Flow 50
anditssuppliedparts.Anynon-observancewillendangeraneective
therapy.
• Withregardtothefundamentalrequirementsofthecurrentlyvaliddirectiveformedicaldevices,the
instructionsforusedescribethepresentstateofthedevice,includingsoftwareandsupplies.
1.3.2)Appropriateapplication
• The
TNI soft Flow 50
isnotintendedforlife-supportmeasures!
• Atherapymayonlybestartedonprescription.
• Onlyhealthcareprofessionalsmayadjusttheprescribedowrate.
• The
TNI soft Flow 50
isasystemdesignedfornasalinsuationtherapy.Itcanbeusedclinically
ininpatientandintensivecareaswellasinhomecaretoproviderespiratoryassistancetospontane-
ouslybreathingpatients.
• Anasalsupplyofrespiratorygasesleadstoapositiveairwaypressure(PAP)dependingonow
rates.ThishastobeconsideredifthePAPcouldcausethepatienttoshowundesirableresults.The
TNI soft Flow 50
isnotintendedforarticiallife-supportmeasures.
• The
TNI soft Flow 50
isnotsuitablefortreatingacutefailureofrespiratoryfunctions.
• The
TNI soft Flow 50
maynotbeusedforinvasiveventilation.
• The
TNI soft Flow 50
maynotbeusedincaseofcompleteclosureoftheupperairways.
• The
TNI soft Flow 50
maynotbeusedonpatientswhoseairwaysarecircumventedbyabypass.
• Duetohumidicationandamorecomfortableaerialapplication,sideeectssuchasanirritationof
thenasalmucosa,blockednose,and(withpatientssueringfromblood-clottingdisorder)bleedings
ofthenose,arequiteunlikelywiththeuseofthe
TNI soft Flow 50
.Ifsuchsymptomsoccur,the
humidityneedstobesettoahigherlevel(seechapter4.1.2“Setting Humidity”).
• Whennotinuse,the
TNI soft Flow 50
shouldnotbeleftswitchedonforseveralhours.
• Donotusethedeviceifyousuerfromepilepticattacksorveryagitatedsleep.
• Donotletchildrenplaywiththehosesorthecablestoeliminatethedangerofstrangulationand
theinhalationorswallowingofsmallparts.
1.3.3)Correctuse
• Theunit’shousingmayonlybeopenedbyauthorizedpersonnel.
• Excludedfromthisisequipmentwhichcanberemovedforusageandcleaningpurposes
withoutusingtools.
• Priortoopening,thedevicehasalwaystobeswitchedoandthesystemhastobediscon-
nectedfromthemains.
• IftheTNI soft
Flow 50isnotbeingusedoveralongperiodoftime,
o switchotheTNI soft
Flow 50,
o disconnectthepowercordfromthesocket,
o removetheapplicatorand
o makesurethatthereisnowaterinthe“humidicationchamberautoll”norinthe
“waterchamberhumidierhomecare”.
• Priortoswitchingon,theTNI soft
Flow 50alwayshastobecheckedforcorrectassembly.
• Checkallpartsfordamagesanddefects.
• Incaseofanyabnormality,switchothepowerswitchanddisconnectthesystemfromthe

Table of Contents

Other manuals for TNI softFlow 50