43
Operating Instruction TNI soft Flow 50 homecare system
Filterclassofdustlter: G4(EN779:2003)
Dustlterchangingcycle: Pleasenote3.1.8
Expected operating time
(Expected Service Life) of TNI soft
Flow 50:
3-6years,dependsondailyusage
Attached Oxygen source
Type: Onlymedicalapprovedoxygensources
maybeconnected(thatincludes,but
isnotlimitedtoIEC60601-1:2005and
forhomecareuseIEC60601-1-11:2010
compliance).Formoreinformation,please
consulttheoxygenusermanualoryour
oxygenretailer.
Forhandlingandadjustment,pleaserefer
totheaccordingusermanual.
Max.pressureallowedattheoxygenintake 200mbar
8) Error codes
SAFETY NOTES
• Some errors lead to the result that the device can no longer be operated.
• In the case of a fault contact your responsible TNI medical AG representative.
• Alarm signals will sound to indicate danger for the user to an error that have occurred.
• The delay between error condition and error signal can be up to one minute max.
• The alarm signals are divided into two steps:
o Alarms with low priority (severity code I):
• 1 audible signal sounds,
• This procedure is repeated cyclically,
• The error code will be indicated in the display,
• When the error is cleared the device is again ready to operate,
• The visual indication in the display cannot be switched o.
o Alarms with medium priority (severity code II):
• 3 audible signals sounds,
• This procedure is repeated cyclically,
• The error code will be indicated in the display,
• These alarms cannot be switched o via menue,
• The device can no longer be operated,
• The user is not able to correct the fault,
• Contact your responsible TNI medical AG representative.