41
Operating Instruction TNI soft Flow 50 homecare system
7.2) Performance parameters, technical data, device parameters
TNI soft Flow 50
PERFORMANCEDATA:
Flowrate 10to50l/min
Adjustablein 0,5l/minsteps,
Tolerances +/-2%ofindexvalue
Admixtureofoxygen 0-20l/minpossible
Withl/min–indicator inthedisplay
FiO
2
indicator integratedinthedisplay
Humiditydewpoint from30°-37°DP,
Adjustablein 1°C-steps
Condensate-freegas-airow forapleasanttherapy
Eventmemory overwiewofthelast12therapymonths
SD-Card storesallevents,mobiledatatransferpossible
individuallanguagememory
Languageselectionfordisplayindicators availableasindividual
standardgermanandenglish
7.3) Technical data, device parameters
Medicalproductclass(93/42/EWG): IIa
Safetyclass,electrically: II
Soundlevel: <30dB(A)
Alarmsignalsoundpressure: >60dB(A)
Safetytype: IP21
Appliedpart(Applicator): BF
Electricalsafety: AccordingtoEN60601-1
UL60601-1
CSAC22.2/No60601-1
Electromagneticcompatibility: AccordingtoEN60601-1-2
Operatingvoltage(nominalvoltage): 100-240VAC,50-60Hz
Maximumpowersystem: 300VA
Maximumpowerheatingplate: 150VA
Maximumpowerapplicatorheating: 48VA