29
Operating Instruction TNI soft Flow 50 homecare system
Settinghumidity(dewpoint)
Theindividualmoisturecontentoftheappliedair/air-oxygenmixturewillbeprovidedviathis
parameter.
• Itisrecommendedtoperformthetherapywith34°-37°dewpointtoensureoptimum
humidicationofthemucosa(mucousmembranes).
• Thehumidity(dewpoint)canbesetfrom30-37°dewpoint.
• Adjustablein1°Csteps.
• The humidity will be set during therapy:
o
eachkeystrokeincreasesthehumidityvalueby1°DP
o eachkeystrokedecreasesthehumidityvalueby1°DP
o
acceptsthevalueandsavesitinthesystem
4.2) Menuitemswhichhasbeenapprovedforsystemconguration
4.2.1)Settingthelanguage
NOTE
Language selection and language diversity
• The TNI soft
Flow 50 basically oers 3 types of languages to select from.
• Additionally, it is possible to store 4 additional languages in the system.
• Additional languages can be selected via SD Card.
SAFETY NOTE
• Concerning availability, please contact your local TNI medical AG agency.
o
o markthe languagevia
o
o choosetherequiredlanguagevia
o
o Endtheadjustmentvia
o Thepreviouslyselectedlanguagenowisthesystemlanguage.