Operating Instruction TNI soft Flow 50 homecare system
12) Electromagnetic compatibility (EMC)
SAFETY NOTE
• The TNI soft Flow 50 is a MEDICAL ELECTRICAL EQUIPMENT and needs special precautions
regarding EMC and needs to be installed and put into service according to the EMC informa-
tion provided in the ACCOMPANYING DOCUMENTS.
• Portable and mobile RF communications equipment can aect MEDICAL ELECTRICAL EQUIP-
MENT!
• The TNI soft
Flow 50 should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary the TNI soft Flow 50 should be observed to verify nor-
mal operation in the conguration in which it will be used.
• The use of other accessories , cables or converters to the device TNI soft
Flow 50 can increase
the emission and reduce the immunity of the TNI soft Flow 50.
• In accordance to the applicable standard the TNI soft
Flow 50 has no essential performance.
• The TNI soft
Flow 50 may be interfered with by other equipment, even if that other equip-
ment complies with CISPR EMISSION requirements.
Guidance and manufacturer’s declaration - electromagnetic emissions
TheTNI soft Flow 50isintendedforuseintheelectromagneticenvironmentspeciedbelow.
ThecustomerortheuseroftheTNI soft Flow 50shouldassurethatitisusedinsuchan
environment.
Emissionstest Compliance Electromagneticenvironment-guid-
ance
RFemissions
CISPR11
Group1
TheTNI soft Flow 50usesRFenergy
onlyforitsinternalfunction.Therefore,
itsRFemissionsareverylowandarenot
likelytocauseanyinterferenceinnearby
electronicequipment.
RFemissions
CISPR11
ClassB
The
TNI soft Flow 50issuitableforuse
inallestablishments,includingdomestic
establishmentsandthosedirectlycon-
nectedtothepubliclow-voltagepower
supplynetworkthatsuppliesbuildings
usedfordomesticpurposes.
Harmonicemissions
IEC61000-3-2
ClassA
Voltageuctuations/
ickeremissions
IEC61000-3-3
Complies
Table 6 -Guidance and manufacturer’s declaration - electromagnetic emissions