Operating Instruction TNI soft Flow 50 homecare system
4) Selecting the therapy parameters and system conguration
menu
WARNING
In order to set the therapy parameters, the device has to be completely assembled and also
needs to be in operating mode as described in chapter 3.
The parameters are called up via the menu and the function keys and also adjusted and
stored (please note capture 3.3).
• Thedevicehasstoredthepreviousparametersettingsandisstartedwiththesesettings
afterarestart.
Thefollowingtherapyparametercanbechanged:
o Humidity ( dew point)in°DP
Thefollowingmenuepointsareadmittedforsystemconguration:
o Language
o Time
o Date
Thefollowingmenuepointssupplysysteminformation:
o System information
o Therapy hours
o Service menue accessblocked,onlyfortechnicalsta!
o Clinic menue accessblocked,onlyfortechnicalsta!
4.1) Therapy parameters
4.1.1)Statementontheindicateddisplayvalues
NominalandRealvalueshownontheoperatingdisplay.
WARNING
NOMINAL and REAL VALUE:
The nominal value describes the value the system should deliver as a regulated therapy value.
Reaching these therapy values and providing them correctly is described as the real value.
In the settings, the nominal value denes the targeted value for the system.
In an optimal situation, the real value diers only slightly from the nominal value.
For more information on tolerance specications see chapter 7 “Technical Data”.
SAFETY NOTE
The real value is mandatory for therapy.
The real value ensures a successful therapy.
Alarms and notes on the display indicate deviations from the tolerance limit.
NOMINALVALUES
• Onthebottomlinethedisplayshowstheprogrammednominalvalues,whichcorresponds
tothelastprogrammedtherapyparameters.
o Humidity (dew point)in°DP
o Flow rate inl/min