Operating Instruction TNI soft Flow 50 homecare system
4) Selecting the therapy parameters and system conguration
menu
WARNING
• In order to set the therapy parameters, the device has to be completely assembled and also
needs to be in operating mode as described in chapter 3.
• The parameters are called up via the menu and the function keys and also adjusted and
stored (please note capture 3.3).
• Thedevicehasstoredthepreviousparametersettingsandisstartedwiththesesettings
afterarestart.
Thefollowingtherapyparametercanbechanged:
o Humidity ( dew point)in°DP
Thefollowingmenuepointsareadmittedforsystemconguration:
o Language
o Time
o Date
Thefollowingmenuepointssupplysysteminformation:
o System information
o Therapy hours
o Service menue accessblocked,onlyfortechnicalsta!
o Clinic menue accessblocked,onlyfortechnicalsta!
4.1) Therapy parameters
4.1.1)Statementontheindicateddisplayvalues
NominalandRealvalueshownontheoperatingdisplay.
WARNING
NOMINAL and REAL VALUE:
• The nominal value describes the value the system should deliver as a regulated therapy value.
• Reaching these therapy values and providing them correctly is described as the real value.
• In the settings, the nominal value denes the targeted value for the system.
• In an optimal situation, the real value diers only slightly from the nominal value.
• For more information on tolerance specications see chapter 7 “Technical Data”.
SAFETY NOTE
• The real value is mandatory for therapy.
• The real value ensures a successful therapy.
• Alarms and notes on the display indicate deviations from the tolerance limit.
NOMINALVALUES
• Onthebottomlinethedisplayshowstheprogrammednominalvalues,whichcorresponds
tothelastprogrammedtherapyparameters.
o Humidity (dew point)in°DP
o Flow rate inl/min