Operating Instruction TNI soft Flow 50 homecare system
Decreasestheparametervalue
Takesthemodiedparametervalue

Closesthemenudisplay
• Allrequiredsettingsareprovidedviathatoperatingprocedure.
• TheTNI soft
Flow 50savesthesesettingsinthedevicememory.
• TheTNI soft
Flow 50restartsthesystemwiththesesettings.
3.4) Attaching the applicator
WARNING
To avoid electric shocks,
be careful not to touch the electric connections of the TNI soft
Flow 50 as soon as the ap-
plicator has been put on.
Hygiene
Due to biocompatibility reasons, the applicator may not continuously be worn for more than
24 hours.
The applicator of the homecare series will be exchanged after 720 therapy hours.
The applicator of the clinic series will be exchanged after 336 therapy hours.
Do not put the heated applicator tube on the device.
Make sure that the applicator is not heated up to more than room temperature (additional
heating e.g. by means of a blanket or radiant heater) since this may cause serious injuries.
Do not use accessories which have not been authorized or recommended by TNI medical AG.
SAFETY NOTE
Position the heated applicator tube away from any electronic monitoring electrode (EEG,
ECG, EMG, etc.) to avoid potential interference with the supervised signal.
Attachingtheapplicator
• Carefullyinsertthenosepinsoftheapplicatorintothenoseandmovethetubeoverthe
earsandplaceitforwardagain.
• Makesurethattheslightlycurvedendswillshowtowardtheface(seeimage1+2).
• Toxtheapplicator,pullthexingsleevetowardsthechin.
• Leavethetubeoftheapplicatorfree,makesurenottojamorbendthetube.
1
2
3
4