3.4) Attaching the applicator
WARNING
To avoid electric shocks,
• be careful not to touch the electric connections of the TNI soft
Flow 50 as soon as the ap-
plicator has been put on.
Hygiene
• Due to biocompatibility reasons, the applicator may not continuously be worn for more than
24 hours.
• The applicator of the homecare series will be exchanged after 720 therapy hours.
• The applicator of the clinic series will be exchanged after 336 therapy hours.
• Do not put the heated applicator tube on the device.
• Make sure that the applicator is not heated up to more than room temperature (additional
heating e.g. by means of a blanket or radiant heater) since this may cause serious injuries.
• Do not use accessories which have not been authorized or recommended by TNI medical AG.
SAFETY NOTE
• Position the heated applicator tube away from any electronic monitoring electrode (EEG,
ECG, EMG, etc.) to avoid potential interference with the supervised signal.
Attachingtheapplicator
• Carefullyinsertthenosepinsoftheapplicatorintothenoseandmovethetubeoverthe
earsandplaceitforwardagain.
• Makesurethattheslightlycurvedendswillshowtowardtheface(seeimage1+2).
• Toxtheapplicator,pullthexingsleevetowardsthechin.
• Leavethetubeoftheapplicatorfree,makesurenottojamorbendthetube.