Operating Instruction TNI soft Flow 50 homecare system
REALVALUESonthedisplay
• Therstlineshowsthehumidity(dewpoint)realvalue
• and
↑
o thesystemheatsup,
• and
↓
o thesystemcoolsdown.
• ThesecondlineshowstheFlowraterealvalue.
o thisvaluecorrespondstothecurrentexact,regulatedowratereleasedinl/min,
o consistingofairorair-oxygenmixture.
• Additionally,theadmixtureofoxygen:
o O
2
admixtureinl/min,
o andtheresultingoxygen(concentration)in%,FiO
2
-value.
4.1.2)Settinghumidity(dewpoint)
WARNING
Humidity and dew point
• In order to set a nominal value, the system requires a setup-time of about <30 minutes.
• If the humidity value is altered (no matter if increasing or decreasing the value), the system
will always require a physical setup-time of about 10 minutes.
• The system is to be used in the recommended environment (see chapter “Technical data”).
SAFETY NOTE
• The Real value is obligatory for therapy.
• If your nose feels dry during therapy,
o check for correct choice and setting of the respective humidier type,
o make sure that the “humidier homecare complete” contains enough water,
o increase humidity.
• If you feel a slight stinging in the nose,
o check for the correct choice and setting of the respective humidier type,
o make sure that the “humidier homecare complete” contains enough water,
o increase humidity.
• If condensate forms in the applicator tube or interfering water droplets spurt out of the
applicator,
o check for correct choice and setting of the respective humidier type,
o check for correct ambient conditions. The room temperature might be too low,
o reduce humidity; humidity might be set too high for the valid ambient conditions.