39
Operating Instruction TNI soft Flow 50 homecare system
Thedustlterapcanberemoveddownwards.
• Removethe“dustlter”.
• Cleanthelterwithbiodegradabledetergents(dishwashingdetergents).
• Carefullyrinsethe“dustlter”withrunningwaterandmakesurethatallcleaningresidues
areremoved.
Itiswashableto30°C.
• Wringthe“dustlter”outanddrywell.
Ensurethatnomoisturecangetintothedevicethroughawetlter.
• Insertadry“dustlter”.
• Insertthedustlterapwithslightpressureuntilitlocks.
Protectthelterfromdirectsunlightandhumidity.
5.1.6)Applicatorinhomecareuse
Cleaningcycle:daily
Theapplicatorisdisinfecteddailybywipingwithbiodegradabledetergents.
Carefullywipethenasalcannuladownwithadampcloth,moistenedwithbiodegradabledeter-
gents.Wipethecompleteapplicatortubeandtheplugdown.
Aftertheperiodofuse,disposetheapplicatorintotherespectivegarbagecontainer(plastic-/do-
mesticwaste).Pleasenotealsochapter11–“Disposal”.
6) Disinfection
Usingandcleaninginstructionsarecorrectlycompliedwithcapture5,aTNI soft Flow 50disin-
fectionisnotnecessary.
Shouldasystemdisinfectionbenecessaryduetoamajorreason,pleasecontactthemanufacturer
oryourresponsibleandauthorisedTNImedicalAGrepresentative.Theyareexclusivelyallowedto
performthisprocedure.