Operating Instruction TNI soft Flow 50 homecare system
 
• Thesystemgoesintooperationandsuppliestherequiredowrateattheapplicator,theair
owisperceptible,thevalueisobviousinthedisplay.
• SupplytheTNI soft
Flow 50withtherequiredoxygen,adjustthevalue.
• Theoxygenappliedisshowninthedisplayinl/min.
• Thesystemmixestheadjustedair-owratewiththeoxygen.
• Theowrateisreferredtoastotalow;i.e.theair-oxygenmixtureissettotheowrate
speciedinthenominalvalue.
• Therealvalueandtheoxygenconcentration,resultingfromit,FiO
2
in%,isshowninthe
display.
3.9) Stopping oxygen supply
SAFETY NOTE
The connection to the oxygen source needs to be set up correctly as described in chapter
3.1.7.
The oxygen supply may only be released with the TNI soft
Flow 50 running.
Do not supply the TNI soft
Flow 50 with oxygen during standby-mode!
Please note chapter 3.1.7.
• Thesystemstillsuppliestherequiredowrateattheapplicator,theairowisperceptible.
• TherealowrateandtheFiO
2
,resultingfromit,isshowninthedisplay.
• Switchotheoxygensupplyattheoxygensource.
• Theoxygenshowninthedisplayrisesto0l/min.
• TheFiO2valueshowninthedisplayindicates21%.

• Thesystemswitchesoff.
• The
TNI soft Flow 50
nowindicatesthestandbydisplay.