Operating Instruction TNI soft Flow 50 homecare system
3.1.5)Insertingtheapplicator
• Choosetherightapplicator(seechapter3.5).Removeitfrom
thepackagingandinserttheapplicatorintothedesignated
openingofthedevice.
Inserttheapplicatorplugfromaboveintotheopeningand
pressdownwithlessforceuntilitsnapsin.
• Thelockinglevermovestotheleftstop.
WARNING
In order to avoid burnings,
do not modify the applicator in any way
make sure that the applicator is not heated up to more than
room temperature (e.g. by means of a blanket, electric re...)
since this may cause serious injury.
do not use insulating sleeves or accessories that have not been
authorized or recommended by TNI medical AG.
SAFETY NOTE
In order to minimize interference with the supervised signal, position the heated applicator
tube away from any electronic monitoring electrode (EEG, ECG, EMG, etc.).
3.1.6)Removingtheapplicator
WARNING
In order to prevent mechanical destruction,
remove the applicator from the retainer without eort.
do not use force if it is not possible to remove the applicator right away.
do not use any tools to remove the applicator.
SAFETY NOTE
Do not let any foreign substances or objects get into the system’s
applicator opening.
Removingtheapplicator
• Movethelockingleverundertheapplicatortotheright.
• Theapplicatorisreleasedfromitslocking.
• Removetheapplicatorplugupwardsfromtheholder.
• Now,thesystemcanbeusedwithanotherapplicator.
3.1.7)Connectingoxygensupply
WARNING
In order to prevent burnings,
do not modify the applicator in any way.
note the safety guidelines in chapter 1.3.6.
do not place the TNI soft
Flow 50 device on the oor. Keep a minimum distance of 40 cm!
keep the minimum distance of 40 cm to the wall.
keep the minimum distance of 80 cm to other electric devices.
note that an external oxygen source not being completely connected may lead to an inadequate
therapy.
Be very careful when handling oxygen! Fire hazard!