46
Operating Instruction TNI soft Flow 50 homecare system
Error category Priority Error
number
Indication in the
display
Description
Motortempera-
tureishigher
than120°C.
II 403 Blowersensor
defective
Blowertemperaturesensor
defective
40 II 404 Fanblowerdefec-
tive
Fanblowerdefective,please
contactyourresponsibleTNI
medicalAGrepresentative
50 II 501 - Displaydefect,pleaserestart
system.Inthecaseoffailure
contactyourresponsibleTNI
medicalAGrepresentative.
Displaydefect.Errorisnotdis-
played,butyoushouldhearan
alarmsignal.
50 II 502 Systemfailure Sensorerrorsonsystem
startup,pleaserestartsystem.
Inthecaseoffailurecontact
yourresponsibleTNImedical
AGrepresentative.
60 II 601 Sensordefective Temperaturesensorapplicator
defective
60 II 602 Airowtoohot Airtemperatureatnasalcan-
nulahigherthan43°C(external
Measurementmandatory)
60 II 603 ext.Measuringele-
mentnotdetected
Externalmeasurementelement
atnasalcannulanotdetected,
pleasecheck.
60 I 651 Applicatorheating
defective
Applicatorheatingdefective.
60 I 652 Applicatornot
detected
Applicatornotdetected,please
change.
60 I 653 Typeofapplicator! Theselectedrealowrateis
toohighforthistypeof
applicator.Reduceowrateor
usealargerapplicator.
70 II 701 Systemfailure EEPROMdamaged,please
restartsystem.Inthecaseof
failurecontactyourresponsible
TNImedicalAGrepresentative.