Operating Instruction TNI soft Flow 50 homecare system
Error category Priority Error
number
Indication in the
display
Description
20 I 255 Dewpointnot
reachable
Selecteddewpointnotreach-
able.
(Dierenceabout1°C,depend-
ingonthesetdewpoint)
30 II 301 Heatingplate
overheated
Heatingplateoverheated.
Pleasecheckhumidier.
Detection:Heatingplatetem-
peratureishotterthanitshould
be(Limitdependingonset
temperature)
30 I 351 Pleaserellwater Waterchamberempty,llup
withwater.
Thewaterlevelisnotmeasured
directly,butcanbedetermined
withtheheatingplate
characteristics.
30 I 352 Heatingplate
defective
Heatingplatedefective(no
power),contactyourresponsi-
bleTNImedicalAGrepresenta-
tive.
30 I 353 Sensordefective Temperaturesensorhumidier
defective
30 I 354 Heatingplate
defective
Heatingplatedefective(nocur-
rent),contactyourresponsible
TNImedicalAGrepresentative.
30 I 355 Sensordefective Systemfailure,pleaserestart
system.Inthecaseoffailure
contactyourresponsibleTNI
medicalAGrepresentative
40 II 401 blowerdefective Blowerblocked,pleaseswitch
systemo/on.Inthecaseof
failurecontactyourresponsible
TNImedicalAGrepresentative
40 II 402 Bloweroverheated Bloweroverheated.
Motortemperatureishigher
than120°C.